Dongguan Hust Tony Instruments Co.,Ltd.

Dongguan Hust Tony Instruments Co.,Ltd. China Manufacturer for Furniture Testing Machines, Textile Testing Equipment, Toys Testing Instrument and universal Tensile Test Machine.

Manufacturer from China
Active Member
10 Years
Home / Products / Flammability Testing Equipment /

ASTM D 2303 Flammability Tester IEC 60587 Standards / High Voltage Tracking Testing Equipment

Contact Now
Dongguan Hust Tony Instruments Co.,Ltd.
City:dongguan
Province/State:guangdong
Country/Region:china
Contact Person:Misslinmei
Contact Now

ASTM D 2303 Flammability Tester IEC 60587 Standards / High Voltage Tracking Testing Equipment

Ask Latest Price
Video Channel
Brand Name :HTI
Model Number :HTB-033
Certification :ASTM D 2303, IEC 60587-1984, GB/T6553-2003
Place of Origin :Guangdong,China(Mainland)
MOQ :1 set
Price :Negotiation
Payment Terms :T/T, L/C, Western Union
Supply Ability :10 sets per month
Delivery Time :10 working days
Packaging Details :strong plywood box package
Power :Electronic
Product Name :Flammability Testing Equipment
Test time :6 hours
Dropping accuracy :0.5%
Weight :200KG
Dimensions :1300L*700W*1800H (mm)
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

ASTM D 2303 IEC 60587 Standards High Voltage Tracking Testing Equipment

Description:

High Voltage tracking tester suit for electronic products, household appliances and other materials resistance to tracking and erosion test, simulate the frequency (48HZ-62HZ) or a corresponding DC Voltage, liquid contaminants and ramps samples by electrical resistance to tracking and erosion measurements assessed under severe environmental conditions of electrical insulating materials resistance to tracking and erosion levels.

Standards:

ASTM D 2303, IEC 60587-1984, GB/T6553-2003

IEC 60587 : Test methods for evaluating resistance to tracking and erosion of electrical insulating materials used under severe ambient conditions

Specifications:

1. The electrode: 0.5mm thick material: stainless steel;

2. The upper and lower electrode distance: 50.0mm ± 0.1;

3. The test voltage: 100V-6000V stepless adjustable;

4. The regulator: output: 0 ~ 250V adjustable, capacity 5KVA;

5. Test transformer capacity: 5KVA, the highest output voltage of AC6000V(or DC6000V optional);

6. High-precision voltage regulator output power: AC220V 6000W,accuracy of ± 1%;

7. The loop current is 60mA, cut off the output voltage;

8. Test time: 6 hours;

9. Dropping accuracy: 0.5%;

10. Voltage stability: + 1%;

11. The mark to judge:

According to standard GB/T6553-2003/4.1.2, the voltage applied to the specimen (2.5k V, 3 5K V and 4.5K V), and with a certain velocity dropsbelow 60mA in the current contamination, persistent 6H but flow is through;

12. Dropping device:

The Leif precision peristaltic pump, flow rate range from 0.00185 to 20 ml / min, the working speed of 0.1 ~ 50 rpm, flow accuracy error: < 0.5%;

13. The output voltage: AC / DC (switchable)

14. The experimental group: five groups

15. Control device: using Siemens PLC+ Taiwan wein view touch screen control, control precision,

simple operation, design humanization, the end of the test automatically generate reports, panel thermal

printer can print test results timely;

16. Studio size: 900L*650W*950H (mm); the volume is 0.56m3.

17. The case: the SUS304 stainless steel (wire drawing);

18. Sewage tank made of 316 stainless steel, with sealing device.

19. The instrument dimensions: 1300L*700W*1800H (mm)

20. Weight: 200KG

Feature:

1. Over-current protection : when the electric current value flowing through the specimen is more than 60mA

2. Selection of series resistance : 1 kΩ, 10 kΩ, 22 kΩ, 33 kΩ 1% Grade 200 W

3. Variable voltage settings : 0 ~ 6 kV

4. 1.5% accuracy about entered value with a voltmeter

5. Using timer, a simple TEST hour setting function by hour / minute / second

6. Notification function : In 6 hours test, if the specimen is bad, the buzzer and flashing light starts working.

7. Syringe pump can create a constant flow conditions (flow rate memory function)

8. Electrical conduction meter can measure pollutants

- Resistivity (0.1Ωm ~ 10 kΩ m) and temperature(0 ~ 100°C) measurements

9. Digital depth gauge : ± 0.01m accuracy

10. Insulated TEST BOX which is safe in high-voltage test

Inquiry Cart 0