Dongguan Hust Tony Instruments Co.,Ltd.

Dongguan Hust Tony Instruments Co.,Ltd. China Manufacturer for Furniture Testing Machines, Textile Testing Equipment, Toys Testing Instrument and universal Tensile Test Machine.

Manufacturer from China
Active Member
10 Years
Home / Products / Flammability Testing Equipment /

Laboratory Flammability Testing Equipment NFPA-701 Methods 2 Textiles and Films Flame Propagation

Contact Now
Dongguan Hust Tony Instruments Co.,Ltd.
City:dongguan
Province/State:guangdong
Country/Region:china
Contact Person:Misslinmei
Contact Now

Laboratory Flammability Testing Equipment NFPA-701 Methods 2 Textiles and Films Flame Propagation

Ask Latest Price
Video Channel
Brand Name :HTI
Model Number :HTB-019
Certification :NFPA 701-2004 Methods 2
Place of Origin :Guangdong,China(Mainland)
MOQ :1 set
Price :Negotiation
Payment Terms :T/T, L/C, Western Union
Supply Ability :15 sets per quarter
Delivery Time :10 working days
Packaging Details :strong plywood box package
Power :Electronic
Product Name :Flammability Testing Equipment
Structure :thick 1.5mm 304 # stainless steel
Furnace specifications :length, width (305 ± 5) mm, height (2130 ± 10) mm
Igniter :high voltage automatic ignition
combustion timing :0-99.99s
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

NFPA-701 Methods 2 Textiles and Films Flame Propagation Testing Equipment

Application:

It is used to assess the propagation of flame beyond the area exposed the ignition source. It is applicable to vinyl-coated fabric blackout linings and fabrics including multilayered fabrics, films and plastic blinds, with or without reinforcement or backing, with area density greater than 700g/m2.

Standard:

NFPA 701-2004 Methods 2

Specification:

1. Laboratory: in accordance with 12.2 of NFPA 701-2004

1.1 Furnace specifications: length, width (305 ± 5) mm, height (2130 ± 10) mm;

1.2 Structure: made of thick 1.5mm 304 # stainless steel. In front of a transparent observation window;

1.3 Top Sample Rod:

1.3.1 from the bottom of the furnace body edge 1190 ± 10mm, both sides should be set to 4mm horizontal alignment of the hole;

1.3.2 consists of steel rods with a diameter of (1.5 ± 0.1) mm or (3.0 ± 0.1) mm and a length of (330 ± 10) mm, which one end is cut into a sharp point;

1.4 Bottom sample rod: consisting of steel rods with a diameter of (1.5 ± 0.1) mm and (255 ± 10) mm in length, which one end shall be cut into a sharp point;

1.5 Vertical guide line:

1.5.1 A pair of vertical guide wires made of soft steel wires are located at the center of the furnace body (100 ± 5) mm on both left and right sides (the distance of the pair of wires is (200 ± 5) mm);

1.5.2 The distance between lines of each pair is (25 ± 2) mm (from front to back in the box)

1.5.3 The vertical guide wire shall be mounted on a rod of diameter (6 ± 0.05) mm

1.6 The upper part of the test room shall be installed with glass fiber cloth baffle with dimensions of (1000 ± 10) mm × (125 ± 5) mm; glass fiber cloth shall be made with glass sewing thread at each end (40 ± 3) mm (3 ± 0.2) mm × (330 ± 5) mm, and pass through the hole (4 ± 0.5) mm in the middle of the top of the chamber; (4 ± 0.5) mm × 25 ± 1) mm, each end is round;

2 Burner:

2.1 Installed in the fixed seat, into a 20-angle tilt;

2.2 Burner heat: 800BTU / ft3 ~ 1000BTU / ft3

2.3 Nozzle from the bottom of the test: 100mm

2.4 Flowmeter: for the rotor flowmeter, the maximum range should be 150 L / hr or 300 L / hr

2.5 Needle Valve: adjust the height of the flame

2.6 Pressure gauge: 0 ~ 1Mpa, the accuracy of 0.5

2.7 Flame measurement scale: measure burner flame height and sample size with an accuracy of 1 mm

3 Control section:

3.1 Igniter: high voltage automatic ignition

3.2 Timers:

3.2.1 combustion timing: 0-99.99s arbitrarily set, the resolution of 0.01s, accuracy of 0.1s

3.2.2 continued combustion time: 0-99.99s arbitrarily set, the resolution of 0.01s, accuracy of 0.1s

3.3: automatic control: such as ignition, combustion timing, continued combustion time.